The NIM interfaces the call points and over door lights with the CarePlus™ Central Control Module (CCM). When multiple NIMs are cascaded via structured cabling, they can be controlled as a single unit allowing any input to link to any output.
Features
16 Inputs (patient stations)
24 Outputs (over door lights)
24V DC
Front and Rear Rack mountable
RS485 Communications (1200, 2400, 4800 and 9600 baud)
Details
1 Input per RJ45 port, consisting of 24V,
Reassurance LED, Call and Cancel contacts
3 Outputs per RJ45 port, each output is common 24V, with 0V switched output
Up to 32 NIMs can be connected together allowing a maximum of 512 inputs and 768 outputs
RS485 over structured cables
Dual Press 3 Press Upgrade
Functionality enables call priority upgrade via 3 press or dual press
Functionality enables nurse presence to auto cancel call
Technical Data
NIM Part Number : NIM
Ambient Temperature : 0 – 50ºC (20-90% RH non condensing)
Storage temperature range : -10 – 60ºC (10-95% RH non condensing)